Precautions for Stainless steel High pressure autoclave reactor

Author:


The Stainless steel High pressure autoclave reactor is to be used at the designated location and according to the instructions for use. Check the test pressure, operating pressure and maximum operating temperature engraved on the main container, and use it within the allowable range of conditions.

The Stainless steel High pressure autoclave reactor is to be used at the designated location and according to the instructions for use. Check the test pressure, operating pressure and maximum operating temperature engraved on the main container, and use it within the allowable range of conditions.

Stainless steel High pressure autoclave reactor
Stainless steel high pressure autoclave reactor is a precision equipment. The sealing ring adopts the form of conical contact sealing, and the purpose of sealing is achieved by tightening the main bolts to press them against each other. Special care must be taken for the sealing cone to avoid damage caused by various collisions. When installing the cover, place the reactor body first, then put the cover in a fixed position and install it on the kettle body carefully. When tightening the main bolts, they must be tightened diagonally and symmetrically several times. Evenly, do not allow the kettle cover to tilt to one side to achieve a good sealing effect, and do not exceed the specified tightening torque to prevent the sealing surface from being squeezed or accelerated wear.
The pressure used by the manometer should preferably be used within 1/2 of the indicated pressure. And often compare the pressure gauge with the standard pressure gauge and make corrections.
The pressure gauge for oxygen should not be mixed with pressure gauges for other gases.
For safety valves and other safety devices, use equipment that has undergone regular inspections and meets the specified requirements.
During operation, care must be taken that the thermometer should be inserted into the reaction solution accurately.
The interior of the stainless steel high pressure autoclave reactor and the pads should be kept clean.
When covering the disc flange cover, the bolts located on the diagonal line should be tightened in turn in pairs.
When the measuring instrument is broken, it is mostly broken on the front and rear sides of its glass surface. Therefore, do not stand in these dangerous places during operation. When danger is expected, remove the glass and replace it with a new one.